Z-0770-2020 Class II Terminated

Recalled by Carl Zeiss Meditec, Inc. — Dublin, CA

FDA device recall Z-0770-2020 was initiated by Carl Zeiss Meditec, Inc. on Oct 16, 2018 and is designated Class II. Reason for recall: Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch. The recall status is terminated (terminated Jun 11, 2021). Affected quantity: 151 devices.

Recall details

Product Type
Devices
Report Date
Jan 8, 2020
Initiation Date
Oct 16, 2018
Termination Date
Jun 11, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
151 devices

Product description

The PLEX Elite 9000, a Swept-Source OCT [SS-OCT] - Product Usage: The PLEX Elite 9000 Swept-Source OCT [SS-OCT] is a non- contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures including, but not limited to, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, optic nerve head, vitreous and choroid. The PLEX Elite SS-OCT angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid.

Reason for recall

Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of CA, TX, NY, MA, OH, PA, IA, , KY, FL, MI, IL, MD and Germany, Japan, Switzerland, France Italy, Australia, Austria, Korea, China, India, Canada, Singapore, Norway, Taiwan, Israel, UAE.

Code information

OBO /HJO Serial Numbers: PE9000-001, PE9000-002, PE9000-003, PE9000-003,PE9000-004, PE9000-005, PE9000-006, PE9000-007, PE9000-008, PE9000-009, PE9000-0010, PE9000-0011, PE9000-0012, PE9000-0013, PE9000-0014, PE9000-0015, PE9000-0016, PE9000-0020, PE9000-0021, PE9000-0022, PE9000-0023, PE9000-0024, PE9000-0025, PE9000-0026, PE9000-0027, PE9000-0029, PE9000-0030, PE9000-0031, PE9000-0032, PE9000-0033, PE9000-0034, PE9000-0035, PE9000-0036, PE9000-0038, PE9000-0039, PE9000-0040, PE9000-0041, PE9000-0042, PE9000-0043, PE9000-0044, PE9000-0045, PE9000-0046, PE9000-0047, PE9000-0048, PE9000-0049, PE9000-0050, PE9000-0051, PE9000-0052, PE9000-0053, PE9000-0054, PE9000-0055, PE9000-0056, PE9000-0057, PE9000-0058, PE9000-0059, PE9000-0060, PE9000-0061, PE9000-0062, PE9000-0063, PE9000-0064, PE9000-0065, PE9000-0066, PE9000-0067, PE9000-0068, PE9000-0069, PE9000-0070, PE9000-0071, PE9000-0072, PE9000-0073, PE9000-0074, PE9000-0075, PE9000-0076, PE9000-0077, PE9000-0078, PE9000-0079, PE9000-0080, PE9000-0081, PE9000-0082, PE9000-0083, PE9000-0084, PE9000-0085, PE9000-0086, PE9000-0087, PE9000-0088, PE9000-0089, PE9000-0090, PE9000-0091, PE9000-0092, PE9000-0093, PE9000-0094, PE9000-0095, PE9000-0096, PE9000-0097, PE9000-0098, PE9000-0099, PE9000-0100, PE9000-0101, PE9000-0102, PE9000-0103, PE9000-0104, PE9000-0105, PE9000-0106, PE9000-0107, PE9000-0108, PE9000-0109, PE9000-0110, PE9000-0111, PE9000-0112, PE9000-0113, PE9000-0114, PE9000-0115, PE9000-0116, PE9000-0117, PE9000-0118, PE9000-0119, PE9000-0120, PE9000-0121, PE9000-0122, PE9000-0123, PE9000-0124, PE9000-0125, PE9000-0126, PE9000-0127, PE9000-0128, PE9000-0129, PE9000-0130, PE9000-0131, PE9000-0132, PE9000-0133, PE9000-0134, PE9000-0135, PE9000-0137, PE9000-0138, PE9000-0139, PE9000-0140, PE9000-0141, PE9000-0142, PE9000-0143, PE9000-0144, PE9000-0145, PE9000-0146, PE9000-0147, PE9000-0148, PE9000-0149, PE9000-0150, PE9000-0151, PE9000-0152, PE9000-0153, PE9000-0154, PE9000-0155, PE9000-0156, PE9000-0158

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0770-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0770-2020. Report a data issue: hello@regdatalab.com.