Z-0775-2014 Class II Terminated

Recalled by Lab Vision Corporation — Fremont, CA

Recall Details

Product Type
Devices
Report Date
January 22, 2014
Initiation Date
December 6, 2013
Termination Date
February 25, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

Estrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision Corporation, Fremont, CA Thermo Fisher Scientific Anatomical Pathology Tudor Rd., Manor Park Rancorn, Cheshire WA7 1TA UK Analyte specific reagent. It is intended for qualitative immunohistochemistry with normal and neoplastic formalin-fixed, paraffin embedded tissue sections, to be viewed by light microscopy.

Reason for Recall

Reports that the product is not performing as expected. In house testing showed no staining on breast cancer tissues.

Distribution Pattern

Worldwide Distribution: US (nationwide) including the states of TX and MA and internationally to: Turkey.

Code Information

PN/ MS-354-P0, LN 354P1206K; Expiration 03/2015.