Z-0775-2017 Class III Terminated
FDA device recall Z-0775-2017 was initiated by Angiodynamics, Inc. on Sep 28, 2016 and is designated Class III. Reason for recall: Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the… The recall status is terminated (terminated Jun 18, 2018). Affected quantity: 996 boxes.
Recall details
- Product Type
- Devices
- Report Date
- Dec 28, 2016
- Initiation Date
- Sep 28, 2016
- Termination Date
- Jun 18, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 996 boxes
Product description
ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO B PG, Item Number H787065970515, Catalog No.REF 06597051, 2) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO S PG, Item Number H787065970535, Catalog No.REF 06597053, 3) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO A PG, Item Number H787065970555, Catalog No.REF 06597055 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick
Reason for recall
Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
Distribution pattern
Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
Code information
Batch/Lot: 1) 5030537, 5037976, 5043444, 5059984, 5066413, 2) 5024545, 5031723, 5038659, 5042534, 5065538, 5071294, 5076428, 5082169, 3) 5022318, 5032456, 5034279, 5081986, 5082446
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0775-2017
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026