Z-0776-2021 Class II Terminated

Recalled by Boston Scientific Corporation — Marlborough, MA

Recall Details

Product Type
Devices
Report Date
January 13, 2021
Initiation Date
December 1, 2020
Termination Date
July 14, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 units

Product Description

Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 10mm x10mm (OUS) UPN: M00553540 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.

Reason for Recall

The rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity

Distribution Pattern

US Nationwide distribution - Foreign: TBD.

Code Information

Lot Number: 26130517 Exp. Date Range: 10/5/2022 GTIN: 08714729904564