Z-0781-2023 Class II Ongoing
FDA device recall Z-0781-2023 was initiated by Steris Corporation on Dec 12, 2022 and is designated Class II. Reason for recall: Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as… The recall status is ongoing. Affected quantity: 105 boxes.
Recall Details
- Product Type
- Devices
- Report Date
- Jan 11, 2023
- Initiation Date
- Dec 12, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 105 boxes
Product Description
VERIFY Steam Test Pack (20 Tests and 20 controls) Item Number: S3069 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes
Reason for Recall
Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.
Distribution Pattern
Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA
Code Information
UDI: 10724995023864 lot #230613