Z-0781-2023 Class II Ongoing

Recalled by Steris Corporation — Mentor, OH

Recall Details

Product Type
Devices
Report Date
January 11, 2023
Initiation Date
December 12, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
105 boxes

Product Description

VERIFY Steam Test Pack (20 Tests and 20 controls) Item Number: S3069 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Reason for Recall

Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

Distribution Pattern

Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA

Code Information

UDI: 10724995023864 lot #230613