Z-0785-2022 Class I Ongoing

Recalled by Philips Respironics, Inc. — Murrysville, PA

FDA device recall Z-0785-2022 was initiated by Philips Respironics, Inc. on Jun 12, 2018 and is designated Class I. Reason for recall: In 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into… The recall status is ongoing. Affected quantity: 229,353 devices.

Recall Details

Product Type
Devices
Report Date
Mar 30, 2022
Initiation Date
Jun 12, 2018
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
229,353 devices

Product Description

Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.

Reason for Recall

In 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into particles. This 2018 correction has been superseded by the device removal and rework described in Philips Respironics 2021 recall action for the same devices.

Distribution Pattern

Global Distribution. US Nationwide.

Code Information

All device serial numbers. Material Numbers 1040001 1054096 1054097 1054260 1054655 1058180 1119867 1040001B 1054096B 1054096NB 1054097B 1054260B 1054655B 1054655TP 1054655TPV 1058180B AU1054096 AU1054096B BR1054096 BT1054096 BT1054260 CA1054096 CA1054096B CN1054096 IT1054096B KO1054096 KR1054096 LA1054096 LG1054096 R1054096 R1054096B R1054097 R1054097B R1054260 R1054260B R1054655 R1054655B R1054655TP R1054655TPV R1119867 RSL1054260 RSL1054260B RTO1054260 RTO1054260B U1054096 U1054096B U1054097 U1054097B U1054260 U1054260B U1054655 RBR1054096 RTO1119867 1032800 1032802 1032804 1040002 1040005 1032800B 1032802B 1032804B 1032804TP 1032804TPV 1040002B 1040005B 1125564B AU1032800 AU1032800B BR1032800 CA1032800 CA1032800B IT1032800B KO1032800 KR1032800 LA1032800 LG1032800 R1032800 R1032802 R1032804 R1032804B R1032804TP R1032804TPV R1040005 R1040005B RTO1040005B SP1032800B U1032800 U1032802 U1032804 U1040005 U1040005B

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0785-2022
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.