Z-0787-2016 Class II Terminated
FDA device recall Z-0787-2016 was initiated by DePuy Orthopaedics, Inc. on Jan 8, 2016 and is designated Class II. Reason for recall: Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin. Surgical delay may result if the punch… The recall status is terminated (terminated Mar 24, 2017). Affected quantity: 351.
Recall Details
- Product Type
- Devices
- Report Date
- Feb 17, 2016
- Initiation Date
- Jan 8, 2016
- Termination Date
- Mar 24, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 351
Product Description
Reason for Recall
Distribution Pattern
Worldwide Distribution-US Nationwide, Armenia, Australia, Austria, Belgium, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, P R China, Portugal, Poland, Russia, Singapore, Slovenia, South Africa, Spain, Sweden, South Korea, Switzerland, Thailand, Turkey, UAE, UK, and Brazil.
Code Information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0787-2016
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab