Z-0803-2015 Class II Terminated
FDA device recall Z-0803-2015 was initiated by Stryker Spine on Oct 6, 2014 and is designated Class II. Reason for recall: Stryker has received a complaint from customers relating to a fracture during the use of the Specialty Audible Torque Wrench in surgery. The recall status is terminated (terminated May 8, 2017). Affected quantity: 15 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 24, 2014
- Initiation Date
- Oct 6, 2014
- Termination Date
- May 8, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 units
Product description
Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to final tighten to 12 Nm the blocker of the XIA 3, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors.
Reason for recall
Stryker has received a complaint from customers relating to a fracture during the use of the Specialty Audible Torque Wrench in surgery.
Distribution pattern
Distributed to GA, IN, MI, NJ, OH, PA, TX & VA.
Code information