Z-0804-2016 Class I Terminated

Recalled by Maquet Cardiovascular Us Sales, Llc — Wayne, NJ

FDA device recall Z-0804-2016 was initiated by Maquet Cardiovascular Us Sales, Llc on Dec 10, 2015 and is designated Class I. Reason for recall: Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the… The recall status is terminated (terminated Jul 6, 2016). Affected quantity: 7 units within US.

Recall details

Product Type
Devices
Report Date
Mar 2, 2016
Initiation Date
Dec 10, 2015
Termination Date
Jul 6, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units within US

Product description

Flow-i Anesthesia System Product Usage: The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment by healthcare professionals trained in inhalation anesthesia administration.

Reason for recall

Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues. Only the Model C20 is affected in the US.

Distribution pattern

US Nationwide in the states of CA, NY, and OR

Code information

Model numbers: C20, C30 and C40 (Only Model # C20 for US) Part numbers: 66 77 200 (US Only), 66 77 300 and 66 77 400 Serial numbers (2743, 2852, 2854, 2855, 2856, 2858

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0804-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0804-2016. Report a data issue: hello@regdatalab.com.