Z-0807-2024 Class III Ongoing

Recalled by Philips North America — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
January 31, 2024
Initiation Date
December 29, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 US; 16 ROW total

Product Description

Achieva 1.5T, Model No. 781178, 781196, 781296

Reason for Recall

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.

Code Information

Model No. 781178 UDI-DI N/A Serial Number 20150; Model No. 781196 UDI-DI N/A Serial Nos. 22270 22095 21113; Model No. 781296 UDI-DI 00884838004108 Serial Nos. 33406 33547 32773 32789 32388