Z-0811-2014 Class II Terminated
Recalled by Ziemer Usa Inc
— Alton, IL
FDA device recall Z-0811-2014 was initiated by Ziemer Usa Inc on Apr 16, 2013 and is designated Class II. Reason for recall: A defect in the controller board can lead to the condenser to the video monitor becoming too warm. This could potentially cause the video monitor to shut off. The recall status is terminated (terminated Dec 27, 2017). Affected quantity: 5 units.
Recall details
- Product Type
- Devices
- Report Date
- Jan 29, 2014
- Initiation Date
- Apr 16, 2013
- Termination Date
- Dec 27, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5 units
Product description
1) The FEMTO LDV Z2 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea at constant depth relative to the corneal surface. 2) The FEMTO LDV Z4 Femtosecond Surgical Laser is an opthalmic surgical laser indicated for use in the creation of the corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface that does not enclose a volume of the cornea. 3) The FEMTO LDV Z6 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants, lamellar keratoplasty or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface that may enclose a volume of the cornea.
Reason for recall
A defect in the controller board can lead to the condenser to the video monitor becoming too warm. This could potentially cause the video monitor to shut off.
Distribution pattern
Worldwide Distribution: US (nationwide) including states of: NJ and IL; and Internationally to: Finland, Japan, and Germany.
Code information
Serial Numbers: FLM4135, FLM3281, FLM2963, FLM3907, FLM3817
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0811-2014
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Ziemer Usa Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works