Z-0811-2018 Class II Terminated
FDA device recall Z-0811-2018 was initiated by NxStage Medical, Inc. on Jan 29, 2018 and is designated Class II. Reason for recall: May contain endotoxin levels which have been confirmed to exceed the ANSI/AAMI quality standard dialysis fluid of 0.5 EU/ml when used to prepare product. The recall status is terminated (terminated Apr 3, 2020). Affected quantity: 16,039 cases/2 units.
Recall details
- Product Type
- Devices
- Report Date
- Mar 7, 2018
- Initiation Date
- Jan 29, 2018
- Termination Date
- Apr 3, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 16,039 cases/2 units
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA (nationwide) and to the countries of : United Kingdom & Spain