Z-0814-2018 Class II Terminated

Recalled by Otto Bock Healthcare Product — Vienna, N/A

FDA device recall Z-0814-2018 was initiated by Otto Bock Healthcare Product on Sep 12, 2017 and is designated Class II. Reason for recall: Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used… The recall status is terminated (terminated Jan 19, 2021). Affected quantity: 75 units.

Recall details

Product Type
Devices
Report Date
Mar 7, 2018
Initiation Date
Sep 12, 2017
Termination Date
Jan 19, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
75 units

Product description

OTTO BOCK Kenevo knee joints Model 3C60

Reason for recall

Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.

Distribution pattern

AK, AR, AZ, CA, CT, DC, GA, IL, IN, KY, LA, MD, MI, MN, NC, ND, NJ, MN, NY, OH, OK, OR, PA, PR, TX, VA, and WA AU, BE, DE, FR, GB, IT, KW, LU, NL, NO, AT, SA, SE, CH, ES, ZA, and HU

Code information

Serial Numbers: 201520007, 201520010, 201520012, 201520008, 201520009, 201522040, 201520002, 201520011, 201520006, 201522046, 201522043, 201520028, 201528134, 201528139, 201520013, 201522045, 201525069, 201525109, 201524051, 201524050, 201549087, 201550012, 201550020, 201550022, 201550017, 201550028, 201550027, 201550016, 201550036, 201550014, 201551005, 201550025, 201551028, 201551030, 201550040, 201551032, 201602031, 201602030, 201603014, 201603016, 201603023, 201602034, 201602036, 201602032, 201551034, 201603029, 201603027, 201603028, 201610008, 201601018, 201610010, 201610015, 201610019, 201610022, 201610020, 201611001, 201611010, 201520027, 201610030, 201610035, 201611012, 201603025, 201611002, 201611018, 201612018, 201612020, 201613021, 201613019, 201615008, 201618017, 201616010, 201613008, 201616004, 201616005, 201616012, 201618023, 201618021, 201618018, 201618019, 201618024, 201617031, and 201618025.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0814-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Otto Bock Healthcare Product? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Otto Bock Healthcare Product: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Otto Bock Healthcare Product →

Source: openFDA record for Z-0814-2018. Report a data issue: hello@regdatalab.com.