Z-0815-2013 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 20, 2013
- Initiation Date
- November 30, 2009
- Termination Date
- September 3, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 209860 electrodes
Product Description
Cloth electrodes with Carbon construction: P/N 201-1231, Cat Nos.: 201-123, 202-454, 203-118, 203-342, 203-454, 203-493, 203-672, 400-877, and 400-877-BJC, 2" x 2" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1241, Cat Nos.: 201-124, 202-455, 203-340, 203-674, and 400-872, 2" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1251, Cat Nos.: 201-138, 202-276, 202-456, 203-336, 203-680, and 400-879, 3" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1381, Cat Nos.: 201-126, 203-678, and 400-856, 2" x 3.5" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1501, Cat Nos.: 201-150 and 400-863, 1.25" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1521, Cat No. 400-852, 3 x 5 Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1541, Cat Nos.:201-154 and 400-865 1.5 x 2.5 Oval Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 Cutaneous electrode.
Reason for Recall
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
Distribution Pattern
Distributed in the states of GA and TN.
Code Information
All Serial Numbers distributed between 11/01/2009 and 01/31/2010