Z-0815-2015 Class II Terminated

Recalled by Philips Ultrasound, Inc. — Bothell, WA

FDA device recall Z-0815-2015 was initiated by Philips Ultrasound, Inc. on Nov 5, 2014 and is designated Class II. Reason for recall: When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection… The recall status is terminated (terminated Aug 15, 2016). Affected quantity: additional 52 international units; 261 units (73 in the US and 188 outside the US); 417 units (289 in the US and 128…

Recall details

Product Type
Devices
Report Date
Dec 24, 2014
Initiation Date
Nov 5, 2014
Termination Date
Aug 15, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
additional 52 international units; 261 units (73 in the US and 188 outside the US); 417 units (289 in the US and 128 outside US)

Product description

Additional devices identified in new FSN MA-FCO 83000190: QLAB versions 10.0 with a2DQ and/or aCMQ plug-ins installed with Philips Xcelera software. *********************************************************************************** Devices identified in the Field Safety Notice ( FSN MA-FCO79500316 ) QLAB Quantification Software. QLAB version 10.0, 10.1 and 10.1.1 with a2DQ and/or aCMQ plug-ins are affected. Catalog # 795041 and Part # 989605315491 Designed to view and quantify image data acquired on Philips Healthcare ultrasound products.

Reason for recall

When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.

Distribution pattern

Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI; and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Germany, Ecuador, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, Norway, New Zealand, Philippines, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Turkey, Taiwan, and Vietnam. ********************************************************************************** Worldwide Distribution -- US, including the states of KY, TN, ME, CA, PA, NC, CT, NY, SC, IL, IN, and WI; and the countries of Argentina, Spain, India, Italy, Austria, Germany, Belgium, Canada, Finland, Switzerland, Greece, Malaysia, Sweden, Australia, Netherlands, Mexico, Peru, Taiwan, Japan, France, and Norway.

Code information

****Additional 52 international units were discovered and added to FCO 79500316** 14D4-N1E2F306-W3BC, 15D5-14041EA3-27F8, 1788-H1DD736A-I3B1, 20B6-13C298A0-2775, 236A-13BD1904-276A, 28C1-13DDA15F-27AB, 3391-13C7A2EF-277F, 3BAD-13B7241B-275E, 4549-N20290C7-G3FB, 464A-1423BC64-2837, 466C-13CCAD3E-2789, 5019-S1EC9257-K3CF, 5581-V1E1931F-U3B9, 5BD9-14234717-2836, 6676-148FDF60-290F, 685C-H1E69D6E-W3C3, 6B10-Q1E11DD2-U3B8, 6C00-142DD102-284B, 7168-1422D1CA-2835, 741C-141D522E-282A, 75CF-J1F6A6F5-M3E3, 7E75-1538H8C3-2A5F, 7EFD-13DBCC2B-27A7, 8719-13CB4D57-2786, 8945-1522HA53-2A33, 8F13-1411DDA9-2813, 937A-P1E5B2D4-K3C1, 9448-1489752A-2902, 972F-14015ED5-27F2, 99E3-13FBDF39-27E7, 9B85-R200BB93-H3F7, 9C97-13F65F9D-27DC, B4C9-141BF247-2827, B65A-P24C5634-R48E, C14C-K1DF489E-I3B4, C25E-13D4ECA8-2799, C7C6-13C9ED70-2783, CC0B-K1F4D1C1-N3DF, D285-13DF7693-27AE, D539-13D9F6F7-27A3, D7CB-142B8681-2846, DFD6-14468F40-287C, E2BD-13BE78EB-276C, E571-13B8F94F-2761, E9B6-Y1E3DDA0-Q3BD, EC6A-S1DE5E04-V3B2, F2D3-13F48A69-27D8, FABC-1466A24E-28BC, FF12-M265FF0C-G4C1, FF45-S1E36853-Q3BC, M1F9-V1DDE8B7-M3B2, and P873-13C88D89-2781. **ADDITIONAL SYSTEM CODES FOR 261 UNITS UNDER FCO 83000190****** 830038; 830069; 830353; 830255 ******************************************************************************************** SERIAL NUMBERS FOR INTERNATIONAL UNITS 1580-14DDC482-29AB, 15A2-1486B55C-28FD, 1777-H208FAFD-P408, 1A2B-T2037B61-I3FD, 1AC6-1529D470-2A43, 1AF9-14A73DB7-293E, 1B3D-13F91F6B-27E2, 2050-14C7C612-297F, 2225-Q24APBB3-K48A, 2326-146B3750-28C6, 25FC-140EA88E-280D, 286C-14B7473E-295E, 2B0F-14DD4F35-29AA, 2B20-14B1C7A2-2953, 2DD4-14AC4806-2948, 2DE5-1480C073-28F1, 2FBA-N203P614-J3FC, 3055-15295F23-2A42, 3088-14A6C86A-293D, 30AA-144FB944-288F, 30CC-13F8AA1E-27E1, 3380-13F32A82-27D6, 3500-S24F1602-P494, 3893-14C1D129-2973, 3B25-151360B3-2A16, 3DFB-14B6D1F1-295D, 409E-14DCD9E8-29A9, 4330-152E6972-2A4C, 4374-14804B26-28F0, 45E4-1528E9D6-2A41, 4617-14A6531D-293C, 4639-144F43F7-288E, 48CB-14A0D381-2931, 48ED-1449C45B-2883, 4AB1-I1F7918F-K3E5, 4B6E-14C6DB78-297D, 4D65-G1F211F3-R3DA, 4E77-13E7B5FD-27BF, 513C-13B6AECE-275D, 538A-14B65CA4-295C, 5903-147FD5D9-28EF, 5AD8-H2021B7A-P3FA, 5BEA-13F7BF84-27DF, 5E9E-13F23FE8-27D4, 60FD-14C6662B-297C, 6130-1443CF72-2877, 62F4-Y1F19CA6-J3D9, 63D3-1469D769-28C3, 6406-13E740B0-27BE, 68E6-15387E10-2A60, 6AFF-V20CA565-U40F, 6B9A-1532FE74-2A55, 6E70-14D66FB2-299C, 6EC5-13FCC9D3-27E9, 7135-14A56883-293A, 718A-13CBC2A4-2787, 76AE-146EE1B8-28CD, 7861-S248367F-S486, 7940-14C07142-2970, 7984-141252F6-2814, 7995-13E6CB63-27BD, 7C27-14385AED-2860, 7EEC-140753BE-27FE, 814B-14DB7A01-29A6, 8342-H206B07C-X403, 8410-14AA72D2-2944, 8421-147EEB3F-28ED, 8454-13FC5486-27E8, 8691-152789EF-2A3E, 86C4-14A4F336-2939, 8C70-13EBD5B2-27C7, 8C81-13C04E1F-2770, 8EAD-1517YB1B-2A1D, 91A5-14636D33-28B6, 9404-15379376-2A5E, 947B-1406DE71-27FD, 9C20-152714A2-2A3D, A4B3-13E5E0C9-27BB, A9C6-14B48770-2958, A9F9-1431F0B7-2853, AEFB-152C1EF1-2A47, AF2E-14A98838-2942, AF50-14527912-2894, B1E2-14A4H89C-2937, B496-149E8900-292C, B77D-141672AB-281C, B930-G1EFC772-R3D5, B9ED-14BF115B-296D, BA42-13E56B7C-27BA, C1D6-15312940-2A51, C1F8-14DA1A1A-29A3, C209-14AE9287-294C, CA25-149E13B3-292B, CA36-14728C20-28D4, CF7C-14BE9C0E-296C, D230-14B91C72-2961, D252-1462VD4C-28B3, D4D3-14DF2469-29AD, D6B9-U235E277-N461, D6CA-R20A5AE4-M40A, D787-14D9A4CD-29A2, D97E-V204DB48-S3FF, DA19-152B3457-2A45, DA90-13FA7F52-27E4, DD00-14A31E02-2935, DD22-144CNEDC-2887, DEE6-R1F9DC10-J3E9, DF81-1520351F-2A2F, E44E-M1EEDCD8-M3D3, E53E-143B9008-2866, EA62-14DEAF1C-29AC, EA73-14B32789-2955, EA84-14879FF6-28FE, EFEC-147CA0BE-28E8, EFFD-1451192B-2891, F01F-13FAQA05-27E3, F510-151FBFD2-2A2E, F707-U24AF64D-S48B, F82A-141512C4-2819, FAEF-13E4KB95-27B7, G5AE-13F994B8-27E3, KG13-14872AA9-28FE, PD75-14C2BBC3-2975, R2A5-14D8BA33-29A1, S81E-14A23368-2934, V537-152A49BD-2A44, Y2D8-1456237A-289C, and YC96-H24A8100-S48B. ********************************************************************************************* SERIAL NUMBERS FOR UNITS IN THE US FOR FCO 79500316 1007-15144B4D-2A18, 1029-14BD3C27-296A, 103A-1491B494-2913, 105C-143AA56E-2865, 12AA-153A5344-2A64, 12CC-14E3441E-29B6, 12FF-1460AD65-28B1, 1332-13DE16AC-27AC, 14A1-U26589BF-Q4C1, 1591-14B23CEF-2954, 15B3-145B2DC9-28A6, 15C4-142FA636-284F, 1812-152F540C-2A4E, 1834-14D844E6-29A0, 1845-14ACBD53-2949, 1856-148135C0-28F2, 1867-1455AE2D-289B, 1878-142A269A-2844, 1B1B-14502E91-2890, 1DAD-14A1BE1B-2933, 1DBE-14763688-28DC, 1DCF-144AAEF5-2885, 1DE0-141F2762-282E, 1E02-13C8183C-2780, 1F71-I24F8B4F-X495, 2083-14452F59-287A, 2247-Q1F2FC8D-H3DC, 2304-14C24676-2974, 2315-1496BEE3-291D, 2337-143FAFBD-286F, 2348-1414282A-2818, 24D9-J2448C17-Q47F, 25B8-14BCC6DA-2969, 25C9-14913F47-2912, 260D-13E320FB-27B6, 261E-13B79968-275F, 285B-14E2CED1-29B5, 287D-148BBFAB-2907, 288E-14603818-28B0, 28B0-140928F2-2802, 2A30-J2651472-Y4C0, 2B53-142F30E9-284E, 2B64-1403A956-27F7, 2B75-13D821C3-27A0, 2DC3-14D7CF99-299F, 2DF6-145538E0-289A, 2E07-1429B14D-2843, 2E18-13FE29BA-27EC, 332B-14CCD061-2989, 334D-1475C13B-28DB, 335E-144A39A8-2884, 35DF-14C750C5-297E, 35F0-149BC932-2927, 37B4-J2499666-K489, 37D6-Y1F28740-S3DB, 38A4-14964996-291C, 38B5-146AC203-28C5, 38E8-13E82B4A-27C0, 38F9-13BCA3B7-2769, 3B47-14BC518D-2968, 3B69-14654267-28BA, 3D2D-N213HF9B-J41C, 40AF-14B15255-2952, 40D1-145A432F-28A4, 40F3-14033409-27F6, 4352-14D75A4C-299E, 4363-14ABD2B9-2947, 4385-1454C393-2899, 43A7-13FDB46D-27EB, 4606-14D1DAB0-2993, 4628-147ACB8A-28E5, 48BA-14CC5B14-2988, 48DC-14754BEE-28DA, 48FE-141E3CC8-282C, 4920-13C72DA2-277E, 4B90-146FCC52-28CF, 4BA1-144444BF-2878, 4D43-K2492119-P488, 4E22-14C15BDC-2972, 4E33-1495D449-291B, 4E44-146A4CB6-28C4, 4E55-143EC523-286D, 4E66-14133D90-2816, 50D6-14BBDC40-2967, 50F8-1464CD1A-28B9, 5109-14394587-2862, 511A-140DBDF4-280B, 52BC-X2129A4E-N41B, 5379-14E1E437-29B3, 562D-14DC649B-29A8, 563E-14B0DD08-2951, 5693-13D73729-279E, 58F2-14AB5D6C-2946, 5947-13D1B78D-2793, 5B95-14D16563-2992, 5BA6-14A5DDD0-293B, 5BB7-147A563D-28E4, 5BC8-144ECEAA-288D, 5BFB-13CC37F1-2788, 5E49-14CBE5C7-2987, 5E7C-14494F0E-2882, 601E-P24E2B68-J492, 610E-149ADE98-2925, 6152-13ECC04C-27C9, 62D2-I248ABCC-X487, 63C2-14955EFC-291A, 6665-14BB66F3-2966, 6687-146457CD-28B8, 6698-1438D03A-2861, 66BA-13E1C114-27B3, 6908-14E16EEA-29B2, 693B-145ED831-28AD, 694C-1433509E-2856, 695D-1407C90B-27FF, 6AEE-K2382CF8-X466, 6BBC-14DBEF4E-29A7, 6E81-14AAE81F-2945, 6E92-147F608C-28EE, 6EA3-1453D8F9-2897, 7067-W201A62D-Q3F9, 7146-1479E0F0-28E3, 7157-144E595D-288C, 73D8-14CB707A-2986, 73FA-14746154-28D8, 740B-1448D9C1-2881, 766A-151DYWX4-2A29, 7962-1469621C-28C2, 7973-143DDA89-286B, 7BD2-1512UICC-2A13, 7E97-14E0F99D-29B1, 7EB9-1489EA77-2903, 7ECA-145E62E4-28AC, 7EDB-1432DB51-2855, 807D-G237B7AB-H465, 815C-14AFF26E-294F, 816D-14846ADB-28F8, 81A0-1401D422-27F3, 81B1-13D64C8F-279C, 8432-145363AC-2896, 86B3-14D07AC9-2990, 86D5-14796BA3-28E2, 8967-14CAFB2D-2985, 8978-149F739A-292E, 8989-1473EC07-28D7, 89CD-13C5CDBB-277B, 8B5E-N1F631A8-H3E2, 8C1B-14C57B91-297A, 8C2C-1499F3FE-2923, 8C3D-146E6C6B-28CC, 8E12-G1F0B20C-W3D7, 8EE0-14947462-2918, 8F02-143D653C-286A, 8F24-13E65616-27BC, 9183-14BA7C59-2964, 91C7-140C5E0D-2808, 9426-14E08450-29B0, 9437-14B4FCBD-2959, 9459-145DED97-28AB, 946A-14326604-2854, 96DA-14DBH4B4-29A5, 96EB-14AF7D21-294E, 998E-14D58518-299A, 99B0-147E75F2-28EC, 99C1-1452EE5F-2895, 9C75-144D6EC3-288A, 9C86-1421E730-2833, 9CA8-13CAD80A-2785, 9E39-U1FB3BF7-N3EC, 9EF6-14CA85E0-2984, 9F07-149EFE4D-292D, 9F18-147376BA-28D6, 9F29-1447EF27-287F, A1BB-14997EB1-2922, A1CC-146DF71E-28CB, A45E-14BF86A8-296E, A491-143CEFEF-2869, A4A2-1411685C-2812, A 712-14BAI70C-2963, A723-148E7F79-290C, A734-1462F7E6-28B5, A745-14377053-285E, A993-15371E29-2A5D, A9B5-14E0TF03-29AF, A9E8-145D784A-28AA, AA0A-14066924-27FC, AC69-14DA8F67-29A4, ACCF-13D561F5-279A, AF1D-14D5YFCB-2999, AF3F-147EPYA5-28EB, AF72-13FB69EC-27E6, B1D1-14CF902F-298E, B1F3-14788109-28E0, B3C8-W1FAC6AA-J3EB, B485-14CA1093-2983, B4A7-1473P16D-28D5, B739-14C490F7-2978, B75B-146D81D1-28CA, B78E-13EAEB18-27C5, BA20-143C7AA2-2868, BCA1-14B991BF-2962, BCC3-14628299-28B4, BCD4-1436FB06-285D, BF22-1536A8DC-2A5C, BF44-14DF99B6-29AE, BF55-14B41223-2957, BF77-145DQ2FD-28A9, BF88-14317B6A-2852, C22B-14578361-289E, C23C-142BFBCE-2847, C24D-1400743B-27F0, C3DE-G230D828-M457, C4DF-1452W3C5-2893, C73E-15262A08-2A3B, C760-14CF1AE2-298D, C782-1478NBBC-28DF, C7A4-1420FC96-2831, C957-X1FA515D-R3EA, C9F2-1520AA6C-2A30, CA14-14C99B46-2982, CCEA-146DSC84-28C9, CEBF-S1EF5225-G3D4, CF9E-14678CE8-28BE, CFC0-14107DC2-2810, CFE2-13B96E9C-2762, D241-148D94DF-290A, D263-143685B9-285C, D4E4-14B39CD6-2956, D4F5-14881543-28FF, D506-145C8DB0-28A8, D517-1431Q61D-2851, D528-14057E8A-27FA, D7BA-1457HE14-289D, D96D-R23062DB-X456, DA4C-14A89D9E-2940, DA6E-14518E78-2892, DCEF-14CEA595-298C, DD11-1477966F-28DE, DFA3-14C925F9-2981, E009-13C3F887-2777, E178-X24B6B9A-V48C, E279-146C9737-28C8, E28A-1441RFA4-2871, E4E9-151535E7-2A19, E50B-14BE26C1-296B, E54F-1410X875-280F, E7E1-146197FF-28B2, E7F2-1436106C-285B, EAA6-143090D0-2850, EAB7-1405Y93D-27F9, EAC8-13D981AA-27A2, ED16-14D92F80-29A1, ED27-14ADA7ED-294A, ED49-145698C7-289C, ED5A-142B1134-2845, ED6B-13FF89A1-27EE, EFCA-14D3AFE4-2996, F00E-14259198-283A, F030-13CE8272-278C, F1C1-P1FEE65F-H3F3, F27E-14CE3048-298B, F2A0-14772122-28DD, F2B1-144B998F-2886, F2E4-13C9S2D6-2781, F532-14C8B0AC-2980, F543-149D2919-2929, F7E6-14C33110-2975, F808-146C21EA-28C7, FAAB-149229E1-2913, FACD-143B1ABB-2865, FB00-13B88402-2760, FD4E-14B831D8-295F, FD5F-148CAA45-2908, FD70-146122B2-28B1, FD81-14359B1F-285A, FFF1-14DE39CF-29AB, HV24-145BA316-28A7, J283-152FC959-2A4F, JDCA-13E915E4-27C2, M559-14D33A97-2996, M58C-1450A3DE-2891, M82F-1476ABD5-28DD, MD97-146BAC9D-28C7, NDA8-1440250A-2870, NM46-140493F0-27F9, R1D7-M234F7DD-R460, SAE3-14712C39-28D2, U2B6-14AD32A0-294A, UB27-13C3NDED-2776, VF4A-T245V164-S480.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0815-2015
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Ultrasound, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Ultrasound, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Ultrasound, Inc. →

Source: openFDA record for Z-0815-2015. Report a data issue: hello@regdatalab.com.