Z-0822-2024 Class II Ongoing
Recalled by American Contract Systems, Inc.
— Temple Terrace, FL
FDA device recall Z-0822-2024 was initiated by American Contract Systems, Inc. on Nov 29, 2023 and is designated Class II. Reason for recall: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating… The recall status is ongoing. Affected quantity: 3196 units.
Recall details
- Product Type
- Devices
- Report Date
- Feb 7, 2024
- Initiation Date
- Nov 29, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3196 units
Product description
Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: a) LAP KIDNEY MODULE, kit number AGKD50L; b) ROBOTIC PROSTATE PACK - 264852, kit number ANPR34S; c) LAP CHOLE - 213388, kit number BFLC02AD; d) GENERAL LAPAROSCOPY PACK, kit number GUGE09I; e) LAP CHOLE PACK, kit number HNLC80I; f) Lap Chole Ovarian Appy, kit number JSLC06I; g) LAP CHOLE, ST LUKES, kit number LMLP08Z; h) ROBOTIC PACK - 242561, kit number MHRB88AJ; i) KIT, GENERAL LAPAROSCOPY, kit number MMLP40F; j) KIT, MAJOR LAPAROTOMY, kit number MMML26F; k) LAP CHOLE - LAP HERNIA PACK - 282807, kit number OWLC07K; l) GENERAL LAPAROSCOPY SAH, kit number SAGL57H; m) ROBOTIC PACK SAH, kit number SARB30K; n) ROBOTIC PACK SAH, kit number SARB30L; o) ROBOTIC PACK SAH, kit number SARB30M; p) LAP CHOLE PACK, kit number SMLC53; q) ROBOTICS PACK, kit number SNRP85D; r) ROBOTIC PACK, kit number SSRO22C; s) ROBOTIC PACK, kit number SSRO22D; t) TPK DAVINCI PROSTATE SYSTEM, kit number TPDV16; u) ASC LAP CHOLE PACK (PS 42279), kit number UICH53Y; v) GENERAL LAPAROSCOPIC PACK, kit number WVGL60E
Reason for recall
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Distribution pattern
US Nationwide distribution.
Code information
a) AGKD50L, UDI/DI 00191072141156, Lot Numbers: 905231, exp. 4/5/2024; 638221, exp. 12/28/2023; 689221, exp. 5/21/2023; b) ANPR34S, UDI/DI 00191072171351, Lot Numbers: 721221, exp. 10/6/2023; 879231, exp. 5/1/2024; 896231, exp. 4/14/2024; 927231, exp. 3/14/2024; 956231, exp. 2/13/2024; 645221, exp. 12/21/2023; c) BFLC02AD, UDI/DI 00191072169075, Lot Numbers: 654221, exp. 12/12/2023; 647221, exp. 12/19/2023; 987231, exp. 1/13/2024; UDI/DI 00191072188830, Lot Numbers: 946231, exp. 2/23/2024; 896231, exp. 4/14/2024; d) GUGE09I, UDI/DI none, Lot Numbers: 768221, exp. 8/20/2023; e) HNLC80I, UDI/DI 00191072146267, Lot Numbers: 797221, exp. 7/22/2023; 679221, exp. 11/17/2023; 995231, exp. 1/25/2024; 911231, exp. 3/30/2024; f) JSLC06I, UDI/DI 00191072126061, Lot Numbers: 839221, exp. 6/10/2023; 766221, exp. 8/22/2023; g) LMLP08Z, UDI/DI none, Lot Numbers: 661221, exp. 12/5/2023; 987231, exp. 1/13/2024; 942231, exp. 2/27/2024; 893231, exp. 4/17/2024; h) MHRB88AJ, UDI/DI none, Lot Numbers: 735221, exp. 9/22/2023; 710221, exp. 10/17/2023; 997231, exp. 1/3/2024; 959231, exp. 2/10/2024; 904231, exp. 4/6/2024; i) MMLP40F, UDI/DI none, Lot Numbers: 933231, exp. 3/8/2024; 879231, exp. 5/1/2024; j) MMML26F, UDI/DI none, Lot Numbers: 989231, exp. 1/11/2024; 896231, exp. 4/14/2024; k) OWLC07K, UDI/DI none, Lot Numbers: 682221, exp. 11/14/2023; 646221, exp. 12/20/2023; 982231, exp. 1/18/2024; 928231, exp. 3/13/2024; l) SAGL57H, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 790221, exp. 7/29/2023; m) SARB30K, UDI/DI none, Lot Numbers: 845221, exp. 6/4/2023; 888221, exp. 4/22/2023; n) SARB30L, UDI/DI none, Lot Numbers: 752221, exp. 9/5/2023; o) SARB30M, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 700221, exp. 10/27/2023; 792221, exp. 7/27/2023; 949231, exp. 2/20/2024; 983231, exp. 1/17/2024; p) SMLC53, UDI/DI 00191072188465, Lot Numbers: 969231, exp. 1/31/2024; q) SNRP85D, UDI/DI none, Lot Numbers: 666221, exp. 11/30/2023; 720221, exp. 10/7/2023; 793221, exp. 7/26/2023; 853221, exp. 5/27/2023; 965231, exp. 2/4/2024; r) SSRO22C, UDI/DI none, Lot Numbers: 891221, exp. 04/19/2023; s) SSRO22D, UDI/DI none, Lot Numbers: 660221, exp. 5/31/2023; 698221, exp. 10/29/2023; 756221, exp. 9/1/2023; 804221, exp. 7/15/2023; 840221, exp. 6/9/2023; 933231, exp. 8/31/2023; 987231, exp. 1/13/2024; t) TPDV16, UDI/DI 00191072128904, Lot Numbers: 856221, exp. 5/24/2023; 744221, exp. 9/13/2023; 675221, exp. 11/21/2023; 932231, exp. 3/9/2024; u) UICH53Y, UDI/DI none, Lot Numbers: 897231, exp. 4/13/2024; 882231, exp. 4/28/2024; v) WVGL60E, UDI/DI 00191072144034, Lot Numbers: 766221, exp. 8/22/2023; 694221, exp. 11/2/2023; 952231, exp. 2/17/2024;
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0822-2024
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for American Contract Systems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for American Contract Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about American Contract Systems, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works