FDA device recall Z-0825-2015 was initiated by DePuy Orthopaedics, Inc. on Nov 7, 2014 and is designated Class II. Reason for recall: Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The… The recall status is terminated (terminated Mar 24, 2015). Affected quantity: 19 units worldwide.

Recall Details

Product Type
Devices
Report Date
Dec 24, 2014
Initiation Date
Nov 7, 2014
Termination Date
Mar 24, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19 units worldwide

Product Description

Ceramic Femoral Head. Intended for use in total hip arthroplasty applications to replace the articular surface of the femoral head in primary hip surgery and for the salvage of a failed previous hip surgery.

Reason for Recall

Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The loosening refers to tissue tension and not the taper connection between the femoral head and stem. The investigations found that the small size sleeve (+1.5 femoral head offset) was incorrectly used in the

Distribution Pattern

Distributed in the states of WV, AZ, and AL, and the countries of Germany, and Finland.

Code Information

Lot 7928614 CATALOG NO. : 1365-36-740 Barcode GTIN: 10603295033684

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0825-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.