Z-0825-2017 Class II Terminated
FDA device recall Z-0825-2017 was initiated by Philips Electronics North America Corporation on Dec 1, 2016 and is designated Class II. Reason for recall: Possibility that a patient or user may be exposed to a very small level of touch current. The recall status is terminated (terminated Aug 31, 2017). Affected quantity: 9.
Recall details
- Product Type
- Devices
- Report Date
- Dec 28, 2016
- Initiation Date
- Dec 1, 2016
- Termination Date
- Aug 31, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 9
Product description
Ingenia MR-OR; 781437, 781438 The MR-OR option is intended for allowing transfer of a neurosurgical patient from OR suite to MR suite and back with tools and instructions for proper positioning and MR imaging of those patients. MR-OR option also facilitates execution of simple neurosurgical procedures in the near proximity of the MR scanner without a need to move the patient back to the OR suite
Reason for recall
Possibility that a patient or user may be exposed to a very small level of touch current.
Distribution pattern
Worldwide Distribution -US including CA, NY and Internationally to United Kingdom, Australia, Norway, and Thailand
Code information
781437: 15001, 15002, 15003 781438: 30001, 30002, 30003, 30004, 30005, 30006