Z-0827-2013 Class II Terminated

Recalled by ConMed Corporation — Utica, NY

FDA device recall Z-0827-2013 was initiated by ConMed Corporation on Jan 28, 2013 and is designated Class II. Reason for recall: Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products. The recall status is terminated (terminated Apr 11, 2017). Affected quantity: Domestic Distribution: 97,450 units; Foreign Distribution: 6,800 units.

Recall details

Product Type
Devices
Report Date
Feb 27, 2013
Initiation Date
Jan 28, 2013
Termination Date
Apr 11, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Domestic Distribution: 97,450 units; Foreign Distribution: 6,800 units

Product description

ELECTROLASE(R) HYFRECATOR(R) ELECTRODES - BLUNT, REF/Catalog Number 7-101-8BX, Disposable Hyfrecator(R) Tips, Sterile, Rx Only --- CONMED CORPORATION USA. Intended to be used as active RF electrodes for desiccation, fulguration and coagulation.

Reason for recall

Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products.

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of China, Dubai and Indonesia.

Code information

Beginning Lot Code 120628X, Ending Lot Code 120917X. [Lot codes on boxes and packaging contain a lot code in the following format: 120917X where 12 represents the year (2012), 09 represents the month (September), 17 represents the day of the month, and X represents the manufacturing shift code (1, 2, 3, 4,5,6 or 9)].

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0827-2013
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for ConMed Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0827-2013. Report a data issue: hello@regdatalab.com.