Z-0827-2022 Class II Ongoing

Recalled by Mevion Medical Systems, Inc. — Littleton, MA

FDA device recall Z-0827-2022 was initiated by Mevion Medical Systems, Inc. on Feb 1, 2022 and is designated Class II. Reason for recall: Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other… The recall status is ongoing. Affected quantity: 39 (36 US, 3 OUS).

Recall details

Product Type
Devices
Report Date
Apr 6, 2022
Initiation Date
Feb 1, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
39 (36 US, 3 OUS)

Product description

Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808)

Reason for recall

Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of FL, MO, NJ, OH, OK, UT, DC and the country of Netherlands.

Code information

All theryapy systems, all hand pendant serial numbers. Therapy system UDI 00864366000100 and UDI 00864366000124.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0827-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Mevion Medical Systems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Mevion Medical Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Mevion Medical Systems, Inc. →

Source: openFDA record for Z-0827-2022. Report a data issue: hello@regdatalab.com.