Z-0827-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 6, 2022
- Initiation Date
- February 1, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 39 (36 US, 3 OUS)
Product Description
Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808)
Reason for Recall
Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of FL, MO, NJ, OH, OK, UT, DC and the country of Netherlands.
Code Information
All theryapy systems, all hand pendant serial numbers. Therapy system UDI 00864366000100 and UDI 00864366000124.