Z-0839-2014 Class II Terminated

Recalled by DeRoyal Industries Inc — New Tazewell, TN

FDA device recall Z-0839-2014 was initiated by DeRoyal Industries Inc on Sep 17, 2013 and is designated Class II. Reason for recall: The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic… The recall status is terminated (terminated Jun 5, 2017). Affected quantity: 25 units.

Recall details

Product Type
Devices
Report Date
Feb 5, 2014
Initiation Date
Sep 17, 2013
Termination Date
Jun 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25 units

Product description

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Foot/Ankle Blanket NS, REF T656NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Reason for recall

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Distribution pattern

Nationwide Distribution.

Code information

Lot Numbers: 32901129, 32980912, 33015541, 33024746, 33105239

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0839-2014
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for DeRoyal Industries Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for DeRoyal Industries Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about DeRoyal Industries Inc →

Source: openFDA record for Z-0839-2014. Report a data issue: hello@regdatalab.com.