Z-0842-2018 Class II Terminated

Recalled by Toshiba American Medical Systems Inc — Tustin, CA

FDA device recall Z-0842-2018 was initiated by Toshiba American Medical Systems Inc on Dec 29, 2017 and is designated Class II. Reason for recall: During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error. The recall status is terminated (terminated Nov 29, 2023). Affected quantity: 17 units.

Recall details

Product Type
Devices
Report Date
Mar 14, 2018
Initiation Date
Dec 29, 2017
Termination Date
Nov 29, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17 units

Product description

INFX-8000F Fluoroscopic X-Ray Systems

Reason for recall

During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.

Distribution pattern

United States

Code information

Serial Numbers: AHA1592004, AGB1482001, AHA1482001, AGA1492002, AHA14Z2002, AGA1552003, UUB13Y2039, AHA15Y2005, AHA1612006, AHB1662007, AGB1682004, AGB1692005, AGB16Z2006, AHB1712008, AHB1712009, AGB1712007, AHA1582003

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0842-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Toshiba American Medical Systems Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Toshiba American Medical Systems Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Toshiba American Medical Systems Inc →

Source: openFDA record for Z-0842-2018. Report a data issue: hello@regdatalab.com.