Z-0849-2021 Class I Terminated

Recalled by Penumbra Inc. — Alameda, CA

Recall Details

Product Type
Devices
Report Date
January 27, 2021
Initiation Date
December 15, 2020
Termination Date
May 14, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27,499 catheters

Product Description

Penumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7KIT The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Reason for Recall

The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.

Distribution Pattern

Worldwide distribution: U.S. (nationwide) including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO , MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. And O.U.S (countries) of: Canada, Chile, Colombia, Costa Rica, Dominican Republic, Hong Kong, Israel, Japan, Korea, Lebanon, Malaysia, Nepal, New Zealand, Oman, Panama, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, United Arab Emirates, and Vietnam.

Code Information

Product Name: Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing Catalog Number: 5MAXJET7KIT UDI Code: 00815948020962 Lot Number Range: C15441 thru C19176 F88882 thru F100278