Z-0850-2024 Class I Ongoing

Recalled by Cardinal Health 200, LLC — Waukegan, IL

FDA device recall Z-0850-2024 was initiated by Cardinal Health 200, LLC on Dec 28, 2023 and is designated Class I. Reason for recall: The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal… The recall status is ongoing. Affected quantity: 397676 units.

Recall details

Product Type
Devices
Report Date
Feb 14, 2024
Initiation Date
Dec 28, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
397676 units

Product description

Cardinal Health Monoject, 60mL Enteral Syringe (purple) - For Use with ENFit Connection System, REF 460SE

Reason for recall

The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal Health recommends that they not be used with enteral syringe feeding pumps.

Distribution pattern

Distribution US nationwide and Canada. Foreign distribution updated 9/6/2024 to only include Canada.

Code information

UDI/DI 26971564466251 (Case), 16971564466254 (Box), 06971564466257 (Each), Lot Numbers: 230501, 230701 and 230702

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0850-2024
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cardinal Health 200, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0850-2024. Report a data issue: hello@regdatalab.com.