Z-0855-2016 Class II Terminated

Recalled by Stryker Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-0855-2016 was initiated by Stryker Howmedica Osteonics Corp. on Dec 23, 2015 and is designated Class II. Reason for recall: The wire packages are correctly marked with a "NON-STERILE" label, however, the enclosed Instructions For Use (IFU) states the device are sterilized via gamma irradiation and should not be… The recall status is terminated (terminated May 19, 2017). Affected quantity: 3,741 units.

Recall details

Product Type
Devices
Report Date
Mar 2, 2016
Initiation Date
Dec 23, 2015
Termination Date
May 19, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,741 units

Product description

Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures.

Reason for recall

The wire packages are correctly marked with a "NON-STERILE" label, however, the enclosed Instructions For Use (IFU) states the device are sterilized via gamma irradiation and should not be resterilized. Since the IFU states the devices are sterile, no instructions for moist heat sterilization are provided.

Distribution pattern

Nationwide Distribution.

Code information

All lots of the following Catalog Numbers: 2704-1-024; 2704-3-018; 6703-1-018; 6703-2-120; 6704-1-018; 6704-2-120 and 6704-3-120.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0855-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0855-2016. Report a data issue: hello@regdatalab.com.