Z-0855-2016 Class II Terminated
FDA device recall Z-0855-2016 was initiated by Stryker Howmedica Osteonics Corp. on Dec 23, 2015 and is designated Class II. Reason for recall: The wire packages are correctly marked with a "NON-STERILE" label, however, the enclosed Instructions For Use (IFU) states the device are sterilized via gamma irradiation and should not be… The recall status is terminated (terminated May 19, 2017). Affected quantity: 3,741 units.
Recall details
- Product Type
- Devices
- Report Date
- Mar 2, 2016
- Initiation Date
- Dec 23, 2015
- Termination Date
- May 19, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,741 units
Product description
Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures.
Reason for recall
The wire packages are correctly marked with a "NON-STERILE" label, however, the enclosed Instructions For Use (IFU) states the device are sterilized via gamma irradiation and should not be resterilized. Since the IFU states the devices are sterile, no instructions for moist heat sterilization are provided.
Distribution pattern
Nationwide Distribution.
Code information
All lots of the following Catalog Numbers: 2704-1-024; 2704-3-018; 6703-1-018; 6703-2-120; 6704-1-018; 6704-2-120 and 6704-3-120.