Z-0857-2016 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

FDA device recall Z-0857-2016 was initiated by Philips Electronics North America Corporation on Nov 25, 2015 and is designated Class II. Reason for recall: Reconstructed ECG leads viewed or printed at the Information Center iX may misrepresent the ECG waveform in specific leads. The recall status is terminated (terminated Sep 6, 2017). Affected quantity: 5569 units.

Recall details

Product Type
Devices
Report Date
Mar 2, 2016
Initiation Date
Nov 25, 2015
Termination Date
Sep 6, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5569 units

Product description

Philips Healthcare PIIC Classic Upgrade, 866117 Physiological, Patient Monitor (With Arrhythmia Detection or Alarm)

Reason for recall

Reconstructed ECG leads viewed or printed at the Information Center iX may misrepresent the ECG waveform in specific leads.

Distribution pattern

Worldwide Distribution-US (nationwide) and the countries of Algeria Australia Austria Belgium Bulgaria Canada Denmark Estonia Finland France Germany Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Kenya Korea Kuwait Luxembourg Malaysia Mongolia Morocco Netherlands New Zealand Nigeria Norway Poland Qatar Romania Russian Federation Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Turkey United Arab Emirates and United Kingdom.

Code information

Software B.O

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0857-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0857-2016. Report a data issue: hello@regdatalab.com.