Z-0861-2019 Class II Ongoing

Recalled by Philips North America, LLC — Andover, MA

FDA device recall Z-0861-2019 was initiated by Philips North America, LLC on Jan 9, 2019 and is designated Class II. Reason for recall: Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users. The recall status is ongoing. Affected quantity: 14.

Recall details

Product Type
Devices
Report Date
Feb 27, 2019
Initiation Date
Jan 9, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14

Product description

PageWriter TC30 w/o trolley Government Bundle, Product 860355, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients

Reason for recall

Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.

Distribution pattern

Domestic distribution nationwide. Foreign distribution worldwide.

Code information

US31510433 US31630351 US31630352 US31630353 US51510712 US51510713 US61510947 US61630668 US61630669 US81731977 US81731978 US91511381 US91511445 USO1511461

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0861-2019
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America, LLC →

Source: openFDA record for Z-0861-2019. Report a data issue: hello@regdatalab.com.