Z-0866-2017 Class II Terminated
Recalled by Northgate Technologies, Inc.
— Elgin, IL
FDA device recall Z-0866-2017 was initiated by Northgate Technologies, Inc. on Nov 21, 2016 and is designated Class II. Reason for recall: Device could cause a patient overpressure situation without any visual or auditory indication or warning The recall status is terminated (terminated Nov 27, 2017). Affected quantity: 49 devices.
Recall details
- Product Type
- Devices
- Report Date
- Dec 28, 2016
- Initiation Date
- Nov 21, 2016
- Termination Date
- Nov 27, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 49 devices
Product description
Insufflator, laparoscopic The ConMed GS2000, Smith & Nephew 500, Novadaq 50L and Nebulae I insufflators are devices that provide CO2 gas distention of surgical cavities for diagnostic and/or operative endoscopy. The insufflator has multiple operating modes which can be used for the following procedures: General Laparoscopic, Pediatric Laparoscopic, Bariatric Laparoscopic and Minimally Invasive Vessel Harvesting procedures
Reason for recall
Device could cause a patient overpressure situation without any visual or auditory indication or warning
Distribution pattern
Worldwide Distribution - US including FL and MA. Internationally to Canada
Code information
Serial Numbers: #89170HXB, #89556DYD, #89652FYA, #89661FYA, #89663FYA, #89664FYA, #89670FYA, #89672FYA, #89675FYA, #89677FYA, #89685FYB, #89697FYB, #89708GYA, #89713GYA, #89719GYB, #89720GYB, #89729GYB, #89732GYB, #89734GYB, #89737GYB, #89889IYD, #90165AZB, #90285BZD, #90387DZB, #90391DZB, #90415DZD, #90438DZD, #90469EZB, #90528EZD, #90555FZB, #90556FZB, #90557FZB, #90558FZB, #90561FZB, #90583GZC, #90621GZB, #90654HZB, #90739IZD, #90869KZD, #90931LZD, #90936LZD, #90954LZD, #90962LZD, #90972LZD, #91105CAD, #91108CAD, #91114CAD, #91120CAD, #91380FAC, #91406FAD, #91471GAC, #91544IAC
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0866-2017
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Northgate Technologies, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works