Z-0869-2017 Class II Terminated

Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-0869-2017 was initiated by Stryker Instruments Div. Of Stryker Corporation on Nov 30, 2016 and is designated Class II. Reason for recall: Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion. The recall status is terminated (terminated Jun 1, 2017). Affected quantity: 1,257 in total.

Recall details

Product Type
Devices
Report Date
Dec 28, 2016
Initiation Date
Nov 30, 2016
Termination Date
Jun 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,257 in total

Product description

8450-107-525 2.5MM Precision Match Head, 13CM, single use, Sterile The Stryker MIS (Minimally Invasive Surgery) product offering, is used to remove osseous material during surgical cases. Each cutting accessory consists of: " A shank (B) with a notch (A), that fits in the attachment on one end and is connected to the head of the bur on the other end. " A head (C) connected to the shank (B) and removes material using machined flutes. The head is machined directly on the shank/neck of the bur.

Reason for recall

Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.

Distribution pattern

Worldwide Distribution - US (nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI and countries of: Canada, Japan, Venlo (Netherlands), France, and UK (United Kingdom).

Code information

Lot Number:16056017, Product Number: 8450-107-525

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0869-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0869-2017. Report a data issue: hello@regdatalab.com.