Z-0875-2015 Class II Terminated

Recalled by Zimmer Manufacturing B.V. — Mercedita, PR

FDA device recall Z-0875-2015 was initiated by Zimmer Manufacturing B.V. on Dec 4, 2014 and is designated Class II. Reason for recall: Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the… The recall status is terminated (terminated Jun 18, 2015). Affected quantity: 68 units.

Recall details

Product Type
Devices
Report Date
Dec 31, 2014
Initiation Date
Dec 4, 2014
Termination Date
Jun 18, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
68 units

Product description

Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

Reason for recall

Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which

Distribution pattern

Worldwide Distribution: US (nationwide) and countries of: France, Germany, Iran, Romania, Serbia, Singapore, Taiwan, and United Kingdom.

Code information

Item No. 00-6200-054-20; Lot 62714305

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0875-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Zimmer Manufacturing B.V.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Zimmer Manufacturing B.V.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Zimmer Manufacturing B.V. →

Source: openFDA record for Z-0875-2015. Report a data issue: hello@regdatalab.com.