Z-0883-2016 Class II Terminated

Recalled by Stryker Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-0883-2016 was initiated by Stryker Howmedica Osteonics Corp. on Jan 21, 2016 and is designated Class II. Reason for recall: During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased. The recall status is terminated (terminated Jun 8, 2017). Affected quantity: 147,838 units in total.

Recall details

Product Type
Devices
Report Date
Mar 2, 2016
Initiation Date
Jan 21, 2016
Termination Date
Jun 8, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
147,838 units in total

Product description

Stryker, T2 K-Wire, 3.2x400mm, Recon Product Usage: Intended for the temporary stabilization of bone segments or fragments.

Reason for recall

During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.

Distribution pattern

US Nationwide Distribution

Code information

Catalog number 1806-3030S; All lots distributed before 10 Nov 2015 with non-expired 5 year shelf life