Z-0902-2025 Class II Ongoing

Recalled by QUIDEL ORTHO — Rochester, NY

FDA device recall Z-0902-2025 was initiated by QUIDEL ORTHO on Dec 5, 2024 and is designated Class II. Reason for recall: Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results. The recall status is ongoing. Affected quantity: 1,541 units (1,125 US, 416 OUS).

Recall details

Product Type
Devices
Report Date
Jan 22, 2025
Initiation Date
Dec 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,541 units (1,125 US, 416 OUS)

Product description

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxin is a widely prescribed cardiac glycoside indicated in the treatment of congestive heart failure and supraventricular arrhythmias. Digoxin measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. One sales unit contains 5 cartridges. Each cartridge holds 18 slides (equivalent to one test), for 90 slides per sales unit. Component: No, not a component

Reason for recall

Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.

Distribution pattern

Domestic: Nationwide Distribution International: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom,

Code information

Model No 8343386; UDI/GTIN 10758750004782; Coating 0272, expiration date; 01-April-2025 The scope of this recall is limited to consignees who received VITROS Chemistry Products DGXN slides in the last 18 months. This will include in-date lots manufactured from coatings (production formulation) 0271 and/or 0272. Of the in date product, the latest expiration date is listed for the coating.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0902-2025
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for QUIDEL ORTHO? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0902-2025. Report a data issue: hello@regdatalab.com.