Z-0923-2017 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

FDA device recall Z-0923-2017 was initiated by Philips Electronics North America Corporation on Dec 1, 2016 and is designated Class II. Reason for recall: Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled. The recall status is terminated (terminated Aug 10, 2017). Affected quantity: 1.

Recall details

Product Type
Devices
Report Date
Feb 1, 2017
Initiation Date
Dec 1, 2016
Termination Date
Aug 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product description

Allura Xper FD10 722026 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electro Physiology(EP).

Reason for recall

Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.

Distribution pattern

US: WA OUS: China, Spain, Korea, Republic of, Germany

Code information

783

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0923-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0923-2017. Report a data issue: hello@regdatalab.com.