Z-0923-2020 Class II Terminated
Recalled by Lumitex Inc
— Strongsville, OH
FDA device recall Z-0923-2020 was initiated by Lumitex Inc on Dec 11, 2019 and is designated Class II. Reason for recall: There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light… The recall status is terminated (terminated Sep 8, 2021). Affected quantity: 6837.
Recall details
- Product Type
- Devices
- Report Date
- Feb 12, 2020
- Initiation Date
- Dec 11, 2019
- Termination Date
- Sep 8, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6837
Product description
Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M1238495, M1238499, M1238503, M1238505, M1238507, M1238508, 2063860-001, 2063862-001, 2063870-001, 2063873-001, 2063874-001, 2071393-001, 2080916-001, 2089424-001
Reason for recall
There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.
Distribution pattern
The products were distributed to the following US states: WI.
Code information
2014 Production Serial Numbers: All Serial numbers beginning with QAAT; 2015 Production Serial Numbers: All Serial numbers beginning with QAAU; 2016 Production Serial Numbers: All Serial numbers beginning with QAAV; 2017 Production Serial Numbers: All Serial numbers beginning with QAAW; 2018 Production Serial Numbers: Serial numbers QAAT60001 through QAAT60156
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0923-2020
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Lumitex Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works