Z-0923-2020 Class II Terminated

Recalled by Lumitex Inc — Strongsville, OH

Recall Details

Product Type
Devices
Report Date
February 12, 2020
Initiation Date
December 11, 2019
Termination Date
September 8, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6837

Product Description

Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M1238495, M1238499, M1238503, M1238505, M1238507, M1238508, 2063860-001, 2063862-001, 2063870-001, 2063873-001, 2063874-001, 2071393-001, 2080916-001, 2089424-001

Reason for Recall

There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.

Distribution Pattern

The products were distributed to the following US states: WI.

Code Information

2014 Production Serial Numbers: All Serial numbers beginning with QAAT; 2015 Production Serial Numbers: All Serial numbers beginning with QAAU; 2016 Production Serial Numbers: All Serial numbers beginning with QAAV; 2017 Production Serial Numbers: All Serial numbers beginning with QAAW; 2018 Production Serial Numbers: Serial numbers QAAT60001 through QAAT60156