Z-0925-2017 Class II Terminated
FDA device recall Z-0925-2017 was initiated by Philips Electronics North America Corporation on Dec 1, 2016 and is designated Class II. Reason for recall: Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled. The recall status is terminated (terminated Aug 10, 2017). Affected quantity: 2.
Recall details
- Product Type
- Devices
- Report Date
- Feb 1, 2017
- Initiation Date
- Dec 1, 2016
- Termination Date
- Aug 10, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2
Product description
Allura Xper FD20/15 R8.2 722058 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electro Physiology(EP).
Reason for recall
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
Distribution pattern
US: WA OUS: China, Spain, Korea, Republic of, Germany
Code information
159, 183
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0925-2017
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab