Z-0926-2014 Class II Terminated
FDA device recall Z-0926-2014 was initiated by ArjoHuntleigh Inc. on Sep 11, 2013 and is designated Class II. Reason for recall: The lock pin can become stuck in the "in" position during the manual rotation, resulting in the inability to prone a patient The recall status is terminated (terminated Jan 8, 2015). Affected quantity: 231 units.
Recall details
- Product Type
- Devices
- Report Date
- Feb 12, 2014
- Initiation Date
- Sep 11, 2013
- Termination Date
- Jan 8, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 231 units
Product description
The Rotoprone therapy System is an advanced patient care system for the treatment and prevention of pulmonary complications. The Rotoprone Therapy System provides up to 62 degrees in both prone and supine positions as well as Trendelenburg/Reverse Trendelenburg positioning. Product Usage: Use of the Rotoprone Therapy System is indicated for patients with severe acute respiratory distress syndrome (ARDS). These are critical patients who are at higher risk of cardiac arrest and mortality. Placing a patient in the prone position may reopen collapsed airways and thus improve pulmonary gas exchange and respiratory mechanics.
Reason for recall
The lock pin can become stuck in the "in" position during the manual rotation, resulting in the inability to prone a patient
Distribution pattern
US Nationwide and Middle East
Code information
Model #209500
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0926-2014
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
Need the full FDA record for ArjoHuntleigh Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.