Z-0927-2015 Class II Terminated

Recalled by Trumpf Medical Systems, Inc. — Charleston, SC

FDA device recall Z-0927-2015 was initiated by Trumpf Medical Systems, Inc. on Dec 3, 2014 and is designated Class II. Reason for recall: Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes The recall status is terminated (terminated Aug 23, 2016). Affected quantity: 3303 total all devices.

Recall details

Product Type
Devices
Report Date
Jan 14, 2015
Initiation Date
Dec 3, 2014
Termination Date
Aug 23, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3303 total all devices

Product description

IR remote control JUX, Material no.1307314, Serial number: to (incl.) 102338310 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0927-2015
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0927-2015. Report a data issue: hello@regdatalab.com.