Z-0928-2024 Class II Ongoing

Recalled by AtriCure, Inc. — Mason, OH

FDA device recall Z-0928-2024 was initiated by AtriCure, Inc. on Nov 28, 2023 and is designated Class II. Reason for recall: AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that… The recall status is ongoing. Affected quantity: 1.

Recall details

Product Type
Devices
Report Date
Feb 7, 2024
Initiation Date
Nov 28, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product description

Packaged, Tank Hose Assembly CMA-US, Product Catalog Number A001056

Reason for recall

AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.

Distribution pattern

US Nationwide. Brazil, Canada, Japan, Taiwan

Code information

UDI-DI: 10840143903076; Lot Number: 129675

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0928-2024
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for AtriCure, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works