Z-0929-2024 Class I Ongoing

Recalled by Philips North America — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
February 14, 2024
Initiation Date
December 15, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
302 US; 264 OUS

Product Description

BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format 2170-3000A 2170-3001A 2170-3002A 2170-3003A

Reason for Recall

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

Distribution Pattern

Domestic distribution nationwide. International distribution worldwide.

Code Information

All serial numbers are affected. No UDI-DI available for this product.