Z-0929-2024 Class I Ongoing

Recalled by Philips North America — Cambridge, MA

FDA device recall Z-0929-2024 was initiated by Philips North America on Dec 15, 2023 and is designated Class I. Reason for recall: Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the… The recall status is ongoing. Affected quantity: 302 US; 264 OUS.

Recall details

Product Type
Devices
Report Date
Feb 14, 2024
Initiation Date
Dec 15, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
302 US; 264 OUS

Product description

BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format 2170-3000A 2170-3001A 2170-3002A 2170-3003A

Reason for recall

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

Distribution pattern

Domestic distribution nationwide. International distribution worldwide.

Code information

All serial numbers are affected. No UDI-DI available for this product.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0929-2024
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America →

Source: openFDA record for Z-0929-2024. Report a data issue: hello@regdatalab.com.