Z-0931-2024 Class I Ongoing

Recalled by Philips North America — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
February 14, 2024
Initiation Date
December 15, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
106 US; 241 OUS

Product Description

BrightView XCT, gamma camera for SPECT; Model Nos.: 6-digit format 882482; 12-digit format 453560462131 453560749161

Reason for Recall

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

Distribution Pattern

Domestic distribution nationwide. International distribution worldwide.

Code Information

All serial numbers are affected. No UDI-DI available for this product.