Z-0931-2024 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 14, 2024
- Initiation Date
- December 15, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 106 US; 241 OUS
Product Description
BrightView XCT, gamma camera for SPECT; Model Nos.: 6-digit format 882482; 12-digit format 453560462131 453560749161
Reason for Recall
Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
Distribution Pattern
Domestic distribution nationwide. International distribution worldwide.
Code Information
All serial numbers are affected. No UDI-DI available for this product.