Z-0933-2013 Class III Terminated
FDA device recall Z-0933-2013 was initiated by Siemens Healthcare Diagnostics, Inc. on Dec 20, 2012 and is designated Class III. Reason for recall: The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding… The recall status is terminated (terminated Aug 26, 2014). Affected quantity: 19, 503 cartons.
Recall details
- Product Type
- Devices
- Report Date
- Mar 20, 2013
- Initiation Date
- Dec 20, 2012
- Termination Date
- Aug 26, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 19, 503 cartons
Product description
Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity in human serum.
Reason for recall
The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).
Distribution pattern
Nationwide Distribution
Code information
Siemens Material Number (SMN) 10444944, Catalog number DF84 -- lot numbers BA3016, EXP 2013-01-16; EC3045, EXP 2013-02-14; EA3073, EXP 2013-03-14; DC3102, EXP 2013-04-12; FC3122, EXP 2013-05-02; EA3157, EXP 2013-06-06; BB3184, EXP 2013-07-03; EC3219, EXP 2013-08-07; EC3220, EXP 2013-08-08; GA3248, 2013-09-05; and FB3290, EXP 2013-10-17