Z-0935-2015 Class II Terminated

Recalled by MedX Holdings, Inc. — Ocala, FL

FDA device recall Z-0935-2015 was initiated by MedX Holdings, Inc. on Dec 3, 2013 and is designated Class II. Reason for recall: Review of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing. The recall status is terminated (terminated Sep 29, 2016). Affected quantity: 9.

Recall details

Product Type
Devices
Report Date
Jan 14, 2015
Initiation Date
Dec 3, 2013
Termination Date
Sep 29, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9

Product description

Lumbar Extension Machine for Physical Therapy

Reason for recall

Review of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing.

Distribution pattern

Worldwide Distribution - USA including MO, IL, ID, and MN and Internationally to Germany and South Korea.

Code information

Serial Numbers: 01531280, 01531279, 01531278, 01531284, 01531283, 01531282, 01531281, 01531288, 01531287

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0935-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for MedX Holdings, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for MedX Holdings, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about MedX Holdings, Inc. →

Source: openFDA record for Z-0935-2015. Report a data issue: hello@regdatalab.com.