Z-0938-2014 Class II Terminated

Recalled by Integra LifeSciences Corp. — Plainsboro, NJ

FDA device recall Z-0938-2014 was initiated by Integra LifeSciences Corp. on Jan 22, 2014 and is designated Class II. Reason for recall: Integra LifeSciences has identified through an investigation of complaints that there may be the potential for the shaft of the Malibu Polyaxial Drivers to break during use. None of the complaints… The recall status is terminated (terminated Dec 17, 2014). Affected quantity: 40.

Recall details

Product Type
Devices
Report Date
Feb 19, 2014
Initiation Date
Jan 22, 2014
Termination Date
Dec 17, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40

Product description

Malibu Polyaxial Screwdriver, 8.0 Plus, Model Number 91-2109; Malibu Polyaxial Screwdriver, Standard, Model Number 91-2110. 510(k): 91-2109 K061342. 91-2110 K051942. The Malibu" System is intended to be used as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur. The System is comprised of single use pedicle screws of varying reduction height, rods, rod connectors, etc. and an accompanying set of reusable surgical instruments. Some of the surgical instruments are specific to the system and others are general surgical instruments. The polyaxial screwdriver is specific to the polyaxial screws in the system.

Reason for recall

Integra LifeSciences has identified through an investigation of complaints that there may be the potential for the shaft of the Malibu Polyaxial Drivers to break during use. None of the complaints reported a patient injury. However, some complaints did indicate the surgery was slightly prolonged while the additional driver provided in the kit was retrieved and used.

Distribution pattern

Worldwide Distribution - USA including IA, FL, LA, IL, CA, TX, AZ, NC, MD, NM, VA, and Internationally to South Africa.

Code information

Model Number(s): 91-2109 (8.0 Plus) 91-2110 (Standard). Lot AL549784F: 91-2109 Malibu Polyaxial Screwdriver 8.0 Plus. Lot AL552404F: 91-2110 Malibu Polyaxial Screwdriver Standard. Lot AL546768F: 91-2110 Malibu Polyaxial Screwdriver Standard. The devices are not distributed as sterile and do not have a expiration date.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0938-2014
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0938-2014. Report a data issue: hello@regdatalab.com.