Z-0939-2022 Class II Terminated

Recalled by Graphic Controls Acquisition Corporation — Buffalo, NY

Recall Details

Product Type
Devices
Report Date
April 20, 2022
Initiation Date
March 15, 2022
Termination Date
July 12, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

CLARAVUE Pre-wired Electrode - IP SET 6V RT ADULT-US Model: 32028780 50603-US

Reason for Recall

Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

Distribution Pattern

US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.

Code Information

All lots UDI: 03700506307319