Z-0941-2026 Class II Ongoing
FDA device recall Z-0941-2026 was initiated by DICOM Grid, Inc. on Nov 18, 2025 and is designated Class II. Reason for recall: Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar… The recall status is ongoing. Affected quantity: 924.
Recall details
- Product Type
- Devices
- Report Date
- Dec 24, 2025
- Initiation Date
- Nov 18, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 924
Product description
Intelerad InteleShare software, with ProViewer component
Reason for recall
Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar reconstruction applied; Manual rotation applied, could result in inaccurate length measurement tool values that could compromise diagnostic accuracy, lead to misdiagnosis or inappropriate follow-up.
Distribution pattern
US: AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ, VA, WA, NC, TN, WI, GA, AK, CA, SD, CO, MA, WV, MN, KS, MD, VT, IA, RI, AL, ID, LA, PR, NH, IN, OK, UT, NE, SC, AR, DE OUS: CA, NZ, AU, GB, BE, FR, IT, NO, SE
Code information
Software versions 3.24.3 - 3.25.2. Date of First Distribution: 2024-08-28. UDI-DI: B228INTELESHARE0