Z-0944-2013 Class II Terminated

Recalled by Smith & Nephew Inc — Memphis, TN

FDA device recall Z-0944-2013 was initiated by Smith & Nephew Inc on Feb 7, 2013 and is designated Class II. Reason for recall: 75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label. The recall status is terminated (terminated Jan 11, 2016). Affected quantity: 90 units.

Recall details

Product Type
Devices
Report Date
Mar 20, 2013
Initiation Date
Feb 7, 2013
Termination Date
Jan 11, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
90 units

Product description

INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 75 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for intramedullary rod fixation.

Reason for recall

75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label.

Distribution pattern

Worldwide Distribution - USA including NM, TN, TX, NC, CA, MI, KY, MA, NJ, OH, and HI. Internationally to Canada, United Arab Emirates, Mexico, Argentina, Japan, and Chile.

Code information

Batch Number 12JM17051

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0944-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Smith & Nephew Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Smith & Nephew Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Smith & Nephew Inc →

Source: openFDA record for Z-0944-2013. Report a data issue: hello@regdatalab.com.