Z-0946-2020 Class I Terminated

Recalled by Philips Respironics, Inc. — Murrysville, PA

FDA device recall Z-0946-2020 was initiated by Philips Respironics, Inc. on Nov 19, 2019 and is designated Class I. Reason for recall: Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This… The recall status is terminated (terminated Mar 31, 2021). Affected quantity: 105.

Recall details

Product Type
Devices
Report Date
Feb 26, 2020
Initiation Date
Nov 19, 2019
Termination Date
Mar 31, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
105

Product description

Garbin EVO Ventilator, Linde - Model Number: LD2110X23B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.

Reason for recall

Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.

Distribution pattern

US: None OUS: CH DE DK IT FR IE GB KW QA PL ES AT PH AE SI CZ

Code information

Model/Material Number: LD2110X23B Serial Numbers: H239876824C6F, H239876835DE6, H23987688E335, H239876943081, H23987698FAED, H239877197F78, H23987731C080, H23987734972D, H240018521FDD, H24001858B087, H2400193657F5, H24739549A36C, H2473957014C5, H247401821EC1, H247406117747, H247407147A36, H2474074261B8, H24746531E009, H247467227EF2, H2474675675DE, H24750031996A, H24750050DC33, H24750144D913, H247522426A37, H2475230432BD, H24752527E54F, H247587538775, H247626140F14, H24801227575A, H2496106842C6, H249611989BD2, H249632560D1E, H24963287ACEF, H249633015DC5, H24970053C182, H2497182710D7, H249719471EDB, H249720843C33, H249724955703, H249891402382, H24999893363B, H251260919873, H25230803E618, H252308867F75, H25230916F2B1, H25230960DA8F, H252309795E96, H25231040699A, H2523105161CB, H252310650D87, H25231134386A, H252311723A3C, H25233073481F, H252330743CA0, H252330752D29, H252330761FB2, H252330770E3B, H25233078F6CC, H25233083CBD7, H25233084BF68, H25233085AEE1, H25233092C386, H25233094A6B0, H25233095B739, H25233097942B, H252331006D50, H252331060866, H252331107488, H25233112579A, H2523311432AC, H252331170037, H25233119E949, H252331205EE0, H252331214F69, H252331227DF2, H25233125094D, H252331263BD6, H252331272A5F, H25233129C321, H252331304738, H2523313156B1, H2523313375A3, H25233139DAF9, H252331466F06, H25235209546F, H25235233D15D, H2523527084A6, H252353009372, H252353108AAA, H25235326C5F4, H252354246AE3, H25235432160D, H2523547360E4, H252356192ED6, H25235633B23C, H25235651C5FE, H25235661EF96, H252356907DD7, H25235714AFEF, H25390190D706, H25390192F414, H253901949122, H2539019580AB, H25390196B230, H25904803A41D,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0946-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Respironics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Respironics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Respironics, Inc. →

Source: openFDA record for Z-0946-2020. Report a data issue: hello@regdatalab.com.