Z-0951-2017 Class II Terminated

Recalled by Natus Manufacturing Limited — GORT, Galway

FDA device recall Z-0951-2017 was initiated by Natus Manufacturing Limited on Oct 15, 2016 and is designated Class II. Reason for recall: The fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting… The recall status is terminated (terminated Jun 24, 2019). Affected quantity: 31.

Recall details

Product Type
Devices
Report Date
Jan 18, 2017
Initiation Date
Oct 15, 2016
Termination Date
Jun 24, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
31

Product description

Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Warmette Blanket Cabinet is designed to store and warm blankets and hospital linens; while, the Natus Warmette Dual Cabinet is designed to store and warm blankets and hospital linens in one compartment and fluids (irrigation and/or injection) in the other compartment within the storage times and warming temperatures recommended by the manufacturers of those items

Reason for recall

The fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting slightly.

Distribution pattern

Worldwide Distribution - US including NJ, NC, TX, IL, VA, NY, FL, CA, MS, WA, IN, and MA. and Internationally to Canada and Costa Rica.

Code information

Warmette Blanket Cabinet Serial Numbers: 10216709, 10216711, 10218559, 10219214, 10219337, 10219338, 10219339, 10219340, 10219387, 10219388, 10219869, 10219870, 10219871, 10219872, 10219873, 1028880. Warmette Dual Cabinet Serial Numbers: 10216713, 10216714, 10218500, 10218501, 10218522, 10218526, 10218528, 10218982, 10219257, 10219389, 10219390, 10220534, 10220536, 10220537, 10220538.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0951-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Natus Manufacturing Limited? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Natus Manufacturing Limited: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Natus Manufacturing Limited →

Source: openFDA record for Z-0951-2017. Report a data issue: hello@regdatalab.com.