Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

FDA device recall Z-0952-2015 was initiated by Medtronic Sofamor Danek USA Inc on Dec 3, 2014 and is designated Class II. Reason for recall: The retaining tabs component of the Set Screw Retaining Drive may be oversized, which will not allow proper mating between the Set Screw Retaining Drive and the set screw. The recall status is terminated (terminated Sep 24, 2015). Affected quantity: 24 units.

Recall details

Product Type
Devices
Report Date
Jan 21, 2015
Initiation Date
Dec 3, 2014
Termination Date
Sep 24, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
24 units

Product description

POWEREASE Instruments Set Screw Breakoff Instrument, REF 2346010, Medtronic Sofamor Danek, USA, Inc. Spinal surgical instrumrent.

Reason for recall

The retaining tabs component of the Set Screw Retaining Drive may be oversized, which will not allow proper mating between the Set Screw Retaining Drive and the set screw.

Distribution pattern

US Distribution to the states of : CA, KS, KY, NC, TX, WA, MI, OH, IA, PA, OR, MS, FL, NY, GA and AR.

Code information

Lot # - CA14C107

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0952-2015
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Sofamor Danek USA Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Sofamor Danek USA Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Sofamor Danek USA Inc →

Source: openFDA record for Z-0952-2015. Report a data issue: hello@regdatalab.com.