Z-0953-2015 Class II Terminated

Recalled by DePuy Orthopaedics, Inc. — Warsaw, IN

FDA device recall Z-0953-2015 was initiated by DePuy Orthopaedics, Inc. on Nov 10, 2014 and is designated Class II. Reason for recall: Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the… The recall status is terminated (terminated Dec 8, 2015). Affected quantity: 7,944.

Recall Details

Product Type
Devices
Report Date
Jan 21, 2015
Initiation Date
Nov 10, 2014
Termination Date
Dec 8, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7,944

Product Description

The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement procedures. The instrument interfaces with several related ATTUNE instruments: keel punch, tibial tower, evaluation bullet, fixed bearing tibial impactor, rotating platform tibial impactor, and femoral impactor, for the purpose of impaction, extraction, and insertion applications at steps within the surgical procedure. Due to the frequency of its use, two impaction handles are supplied in each kit.

Reason for Recall

Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.

Distribution Pattern

Worldwide Distribution-US (nationwide) except AK and RI, including PR and the countries of Chile, Canada, Australia, Austria, Belgium, France, Hong Kong, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK

Code Information

CATALOG NO. : 2544-01-017 LOT NUMBER: All Lots Barcode GTIN: 10603295423324

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0953-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.